GE Kaye LTR -40/140 Drywell Low Temperature Reference Dry Well Block -40...140°C

Artikelnummer: 2934

Kategorie: Block Kalibratoren


2.975,00 €

inkl. 19% USt. , zzgl. Versand (Paket)

momentan nicht verfügbar


Manufacturer: GE Kaye

Model: LTR-40/140 Drywell Temperature Reference

S/N: A6C289H

Description:

The LTR-40/140 Drywell temperature reference is specifically designed to fully automate sensor calibration when used with validation software and traceable temperature standards. The LTR-40/140 employs unique inserts which minimize cooling of the thermocouple tips due to heat conduction providing an uncertainty of merely 0.1°C. Without proper inserts, for example, transfer uncertainty in excess of 0.5°C can occur with 22 gage, type T thermocouples. The LTR-40/140 features fast heat up and cool down, a capacity to hold up to 18 to thermocouples and require no messy oils. Calibrating sensors in the environment where they will be used provides the highest degree of accuracy as system temperature coefficient errors are eliminated. The references are lightweight and rugged so you can calibrate sensors almost anywhere.

Features:

- Fast heat up and cool down
- Dry block requires no messy oils
- Capacity to hold from of up to 18 thermocouples
- Software interface with Kaye Validation systems for automatic sensor calibration

For applications ranging from -40°C to 140°C such as:

- Autoclave/ Steam Sterilizers
- Washer Disinfector
- Fermenter
- Steam-in-Place
- Freezer/ Fridge
- Cold Rooms
- Stability Chambers
- Incubators
- Warehouse
- Glove Box/ Incubator

Specifications:

- Temperature Range: -40°C to 140°C
- Ambient Operating Range: 5°C to 50°C
- Set Point Accuracy: 0.2°C
- Temperature Stability: 0.02°C
- Transfer Calibration Accuracy IRTD Standard to Thermocouples:
-40°C to –25°C: ±0.15°C
-25°C to 80°C: ±0.1°C
80°C to 130°C: ±0.15°C
130°C to 140°C: ±0.18°C
- Typical Heat-Up Time:
25°C to 80°C: 6 minutes
25°C to 140°C: 14 minutes
- Typical Cool-Down Time:
10 minutes: 25°C to –20°C
- Access Opening / Well configuration:
Fixed Insert
Reference wells (2): 6.7 diameter x 155 mm deep
Calibration wells (6): 9 diameter x 155 mm deep
- Display: LED w/0.01°C resolution
- Computer Interface: RS232
- Dimensions: 343 x 198 x 318
- Power: 230 VAC 50 Hz, 1.5 A 350 watts

Items delivered:

- GE Kaye LTR-40/140 Drywell Temperature Reference

Condition:

The whole item is in perfect cosmetic as well as technical condition. It runs as intended being 100% functional tested. It is ready for you to use!


Versandgewicht: 30,00 Kg

Bitte melden Sie sich an, um einen Tag hinzuzufügen.